Table 7 Results of method comparison: Total observed error

From: Evaluation of a novel quantitative canine species-specific point-of-care assay for C-reactive protein

TEobs at three CRP concentration ranges / analyzer

ABX Pentra 400 vs POCT

AU 5800 vs POCT

ABX Pentra 400 vs AU 5800

≥ 10 - < 50 mg/l

27.3%

17.3%

10.7%

≥ 50 - < 100 mg/l

26.7%

16.7%

12.6%

≥ 100 - < 200 mg/l

31.9%

21.9%

12.2%

  1. Total observed error (TEobs) for the CRP analysis performed on two automated bench top analyzers (ABX Pentra 400, AU 5800) and the point-of-care analyzer TEobs > TEdes (i.e., 29.6%) but < TEmin (i.e., 44.4%) is shown in bold letters
  2. Abbreviations: vs versus, TE des desirable total allowable error, TE min minimally acceptable total error