Skip to main content

Table 2 Trending ability with concordance and polar plot data for 18 anaesthetized rabbits

From: Prospective clinical study to evaluate an oscillometric blood pressure monitor in pet rabbits

 

Systolic

Mean

Diastolic

Concordance rate

 Exclusion zone 5%

67%

84%

67%

 Exclusion zone 10%

65%

87%

72%

 Polar plot analysis (n)

82

100

90

Mean polar angle

 Exclusion zone 5%

−13.7°

−9.5°

− 12.0°

 Standard deviation

28.0°

28.1°

31.4°

Radial limits of agreement

 Exclusion zone 5%

− 56.4° to 42.2°

−65.1° to 46.6°

−70.2° to 50.8°

  1. Data points representing percentage changes lower than 5% (exclusion zone 5%) or 10% (exclusion zone 10%) were not included in the analysis. n: number of observations